Following the reported event, a steris service technician spoke with facility personnel who stated that contrary to the reported event, the employee did not obtain a cut.No injury occurred and no medical treatment was sought or administered.The technician arrived onsite inspect the 3085 surgical table that was utilized with the armboard subject of the event; however, facility personnel could not confirm which of their ten 3085 surgical tables was utilized at the time of the reported event.The technician inspected the side rails of all ten surgical tables where the armboard connects to the table and the tables were found to be operating properly.No repairs were required.The armboard subject of the event was evaluated by steris engineering and found to be operating properly.No sharp edges were identified.The customer was provided a replacement armboard.No additional issues have been reported.
|