| Model Number |
MMT-751PNAS |
| Medical Device Problem Codes |
Mechanical Problem (1384); Device Difficult to Program or Calibrate (1496); Battery Problem (2885)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
11/06/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.No conclusion can be drawn at this time.We therefore consider this report complete to the best of our knowledge.
|
| |
|
Event or Problem Description
|
|
The customer reported via phone call that they insulin pump button was less responsive.Customer¿s blood glucose level was unknown.Customer reported that the insulin pump diminished battery life, battery not lasting 7 days and slower and louder in rewind.The insulin pump will not be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Device received with no button response due to corroded keypad traces.No button error alarm during testing noted.Unable to perform all functional testing including the displacement, operating currents, a21 error test, rewind, basic occlusion, occlusion, prime a33 and excessive no delivery test or verify rewind anomaly and change/battery lasts less than expected anomaly due to buttons not responding.
|
| |
|