| Model Number |
MMT-523NAP |
| Medical Device Problem Codes |
Device Difficult to Program or Calibrate (1496); Patient Device Interaction Problem (4001)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
10/15/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).A complete analysis and testing of the insulin pump showed that it was functioning properly and passed all functional testing.After testing it was concluded that the device operated within specifications.Device passed the displacement test and self test.All buttons functioning properly.No damage in the keypad assembly noted.The keypad connector j2/lcd locked properly during visual inspection.No rainbow effect on the screen noted during visual inspection.The asr exemption e2015043 was revoked on february 4, 2019.This event was previously reported as an asr.Additional information or analysis results regarding this event will be reported as an mdr.
|
| |
|
Event or Problem Description
|
|
The customer's family reported via phone that the esc button was hard to press.Customer¿s blood glucose was unknown at the time of incident.The insulin pump will be returned for analysis.
|
| |
|