|
As reported by the physician, the 6f.070 3.5 100cm vista brite tip guiding catheter (gc) was fractured out of the box when it was passed to the table.Another catheter was therefore used, and the procedure was performed.The patient came out in perfect condition.There was no reported patient injury.The device was stored as per the labeling and was prepped as per instruction of use (ifu).There was no visible damaged noted to the device prior to opening the packaging.There was an unusual forced used.There were no images available for evaluation.
|
|
After further review of additional information received the following sections d10, g4, g7, h2, h3 and h6 have been updated accordingly.The 6f.070 3.5 100cm vista brite tip guiding catheter was fractured out of the box when it was passed to the table.Another catheter was therefore used, and the procedure was performed.The patient came out in perfect condition.There was no reported patient injury.The device was stored as per labeling.The device was prepped as per instruction of use (ifu).There was no visible damage noted to the device prior to opening the packaging.There was an unusual forced used.There were no images available for evaluation.A non- sterile unit of a vista brite tip guiding catheter (6f.070 xb 3.5 100cm) was received coiled inside of a plastic bag.Several kinks were observed on the body of the catheter.Damages are located at 8.5, 13.5, 43.5, 49, 54, 64, 67, 96 and 97cm from the distal tip.The reported fracture condition was not noticed on the returned device.No other damages or anomalies on the returned unit were observed.Dimensional analysis was performed to verify the correct catheter inner diameter (id) and outer diameter (od).Measurements were taken at three places.Dimensional analysis results were found within specification.A product history record (phr) review of lot 17853778 revealed no anomalies or non-conformances during the manufacturing and inspection processes that can be associated with the reported event.The failure reported by the customer as ¿catheter (body/shaft) ¿ cracked - during prep" was not confirmed since the reported fracture condition was not noticed on the returned device.However, several kinks were observed on the body of the catheter.Exact cause of these damages on the catheter could not be conclusively determined during the analysis.Dimensional analysis results were found within specification and do not suggest that the reported event could be related to the manufacturing process.Storage/handling factors might have contributed to this issue.Per the instructions for use (ifu), which is not intended as a mitigation of risk, ¿store in a cool, dark, dry place.Do not use open or damaged packages.Do not resterilize.Do not expose to organic solvents.Inspect the guiding catheter before use to verify that its size, shape, and condition are suitable for the specific procedure.¿ the recommended procedure section in the ifu states, ¿remove the guiding catheter from its packaging.Inspect the guiding catheter upon removal from packaging to verify that it is undamaged.Warning: do not use a guiding catheter that has been damaged in any way.If damage is detected, replace with an undamaged guiding catheter.¿ neither the phr review nor the product analysis suggests that the event experienced by the customer could be related to the manufacturing process; therefore, no corrective/preventive action will be taken at this time.
|