Section d10, h3, h4: additional information.Manufacturer's investigation conclusions: the reported event of the console not displaying the flow parameter during the maintenance check was not confirmed.The centrimag 1st generation primary console (serial #: (b)(6) was returned for analysis and was evaluated.The console returned with a damaged lcd.The 1st generation centrimag consoles are unable to be repaired and the console was subsequently scrapped, and no further testing was performed.The root cause for the reported event was not conclusively determined through this analysis; however, the observed damage to the lcd may have contributed to the reported event.No further information was provided.The manufacturer is closing the file on this event.
|