• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEXCOM, INC. DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEXCOM, INC. DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM Back to Search Results
Model Number MT22608
Medical Device Problem Code No Audible Alarm (1019)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 10/20/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
It was reported that the receiver had no audio output.No product or data was provided for evaluation.Confirmation of the allegation and a probable cause could not be determined.No injury or medical intervention was reported.
 
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
The product was evaluated.An exterior visual inspection was performed and passed.Charged and boot were performed and passed.Receiver functional testing was performed and failed due to failed audio.Opened receiver case for interior inspection was performed and passed.Speaker resistance was measured and failed.The receiver log was downloaded and reviewed finding no errors related to the complaint.The allegation was confirmed.The probable cause was determined to be a defective speaker.No injury or medical intervention was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM
Common Device Name
CONTINUOUS GLUCOSE MONITOR
Manufacturer (Section D)
DEXCOM, INC.
6340 sequence dr.
san diego CA 92121
MDR Report Key9304521
Report Number3004753838-2019-084317
Device Sequence Number2195776
Product Code MDS
UDI-Device Identifier10386270000184
UDI-Public10386270000184
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
P120005
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/30/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberMT22608
Device Catalogue NumberSTK-PR-001
Device Lot Number5202421
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/20/2019
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 10/20/2019
Supplement Date Received by Manufacturer12/05/2019
Initial Report FDA Received Date11/11/2019
Supplement Report FDA Received Date12/30/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age66 YR
Patient Weight82
-
-