| Model Number |
MT22608 |
| Medical Device Problem Code |
No Audible Alarm (1019)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
10/20/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that the receiver had no audio output.No product or data was provided for evaluation.Confirmation of the allegation and a probable cause could not be determined.No injury or medical intervention was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
The product was evaluated.An exterior visual inspection was performed and passed.Charged and boot were performed and passed.Receiver functional testing was performed and failed due to failed audio.Opened receiver case for interior inspection was performed and passed.Speaker resistance was measured and failed.The receiver log was downloaded and reviewed finding no errors related to the complaint.The allegation was confirmed.The probable cause was determined to be a defective speaker.No injury or medical intervention was reported.
|
| |
|