| Model Number |
22011 M |
| Medical Device Problem Code |
Appropriate Device Problem Term/Code Not Available (3191)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
04/22/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was that the shunt failed to keep necessary pressure.The shunt was implanted on the right side in the kocher's point for combination dysfunction complex, occlusive hydrocephalus.The shunt was explanted and replaced with another device.
|
| |
|
Additional Manufacturer Narrative
|
|
The returned device was patent.The valve met the requirements for leak and reflux testing.The valve did not meet the requirements of pressure flow and preimplantation testing.Biological debris was observed on the interior of the valve.The instructions for use cautions, ¿shunt obstruction may occur in any of the components of the shunt system.The system may become internally occluded due to tissue fragments, blood clots, tumor cell aggregates, bacterial colonization or other debris.¿ all valves are 100% tested at the time of manufacture.If information is provided in the future, a supplemental report will be issued.
|
| |
|