| Model Number |
A140-SV |
| Medical Device Problem Codes |
Detachment of Device or Device Component (2907); Device Damaged by Another Device (2915); Material Twisted/Bent (2981); Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
10/23/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
During atherectomy of the tibial-peroneal trunk with the pantheris sv catheter, imaging was lost during advancement of the catheter.The physician noted that he had applied additional force just before imaging was lost.During retraction of the catheter, the nosecone detached from the cannula while the cannula and nosecone were within the sheath and both were withdrawn from the patient.No part of the device was left inside the patient.The sheath and guidewire did not have to be removed.The procedure continued with another pantheris sv device advanced through the sheath and was completed with no reports of patient injury or harm.The ifu for this device states as a precaution for the user to determine the cause of resistance to movement of the catheter before continuing to move the device against resistance.
|
| |
|