|
H10: manufacturing review: the lot number was provided, and the lot device history records have been reviewed.The lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: based on the investigation of the returned delivery system a tip detachment was confirmed.Upon investigation the deployment mechanism was found in used condition because the stent had been deployed inside patient; the inner catheter tip was missing.A definite root cause for the reported event could not be determined.Labeling review: based on the lot number reported, the instructions for use (ifu) were supplied with the product.They are currently valid, and therefore relevant for review: in reviewing the relevant labeling for this product it was found that the instructions for use (ifu) sufficiently addressed the potential risks.Regarding preparation the ifu state: 'flush the inner lumen of the device with saline prior to use.' regarding accessories the packaging labels indicate the use of a 6f introducer and a.035" guidewire.The ifu further states: if excessive force is felt during stent deployment, do not force the stent system.Remove the stent system as possible, and replace with a new unit.Furthermore, the ifu state: 'examine the stent system for any damage.If it is suspected that the sterility or performance of the stent system has been compromised, the device should not be used.' regarding predilation the ifu states: 'predilation of the lesion should be performed using standard techniques.'.H10: d4(expiry date: 09/2020), g4.H11: h3, h6(results, conclusion), g3.H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|