Upon receipt, the lead under complaint was found cut approx.17 cm distal to the is-1 connector pin.Only the proximal fragment was received for analysis.The distal fragment including the lead tip was not received for analysis.The returned lead fragment was visually and electrically analyzed.The visual inspection showed that the is-1 connector pin was sheard off.In addition, deformations of the inner coil close to the is-1 connector pin were found which most likely were caused by overrotation of the inner coil during the retraction of the fixation helix.Further investigation revealed that the insulation was, apart from the present fragmentation, free of breaches.During the electrical analysis, the dc resistances of the received conductor fragment were measured.No peculiarities were found.Further analysis of the returned lead fragment did not reveal any irregularity that might have contributed to the reported clinical observation.The manufacturing process for this lead was re-investigated.All production steps had been performed accordingly.There was no sign of any inconsistency during the manufacturing process.In conclusion, the analysis did not reveal any sign of a material or manufacturing problem.
|