Model Number LF1723
Medical Device Problem Code Energy Output Problem (1431)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 09/27/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, intra operatively, the quality of the sealing of the device decreased after 10 minutes of use.They used a new device to finish the procedure.There was no patient injury.
 
Additional Manufacturer Narrative
Additional information: evaluation summary: one device was received for evaluation.The returned product did not meet specification as received.Initial visual inspection found no defects.The reported condition was confirmed.The investigation found that regrasp alarms sounded continuously and the device could not be activated.An end tone and seal cycle could not be completed.The device was disassembled, and investigation found the gray wire near the jaw was damaged.The investigation identified the root cause of the reported event to be a break in the wire that occurred from a kink.The break got larger as the device was cycled during manufacturing and the procedure, until it shorted against the inner tube guide and caused a re-grasp alarm.The kink from the wire likely occurred when the jaw molding supplier was packing the jaws for shipment.Corrective actions have been implemented to mitigate this issue.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIGASURE
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
COVIDIEN MFG DC BOULDER
5920 longbow dr
boulder CO 80301 3299
MDR Report Key9307214
Report Number1717344-2019-01491
Device Sequence Number11374595
Product Code GEI
UDI-Device Identifier20884521181585
UDI-Public20884521181585
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
K141153
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,u
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 12/23/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/14/2022
Device Model NumberLF1723
Device Catalogue NumberLF1723
Device Lot Number73190199X
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/22/2019
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/25/2019
Supplement Date Received by Manufacturer11/27/2019
Initial Report FDA Received Date11/11/2019
Supplement Report FDA Received Date12/23/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-