NIHON KOHDEN CORPORATION AG-920RA; MULTI-GAS UNIT Back to Search Results
Model Number AG-920RA
Medical Device Problem Codes Gas Output Problem (1266); No Device Output (1435)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 10/17/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The customer reported that the multi-gas unit failed during patient monitoring.The device displayed error messages of, "check water trap and sample line" and "check external unit." no patient harm was reported.The device has been returned to nihon kohden and is awaiting evaluation.Nihon kohden continues to investigate the reported event.Nihon kohden will submit a supplemental report in accordance with 21 cfr section 803.56 when additional information becomes available.
 
Event or Problem Description
The customer reported that the multi-gas unit failed during patient monitoring.The device displayed error messages such as, "check water trap and sample line" and "check external unit." no patient harm was reported.
 
Event or Problem Description
The customer reported that the multi-gas unit failed during patient monitoring.The device displayed error messages such as, "check water trap and sample line" and "check external unit." no patient harm was reported.
 
Additional Manufacturer Narrative
Details of complaint: on 10/17/2019 customer stated that gas device was displaying error message for checking water trap and sampling line, even after replacing these parts.Device was sent to nka for evaluation and repair.Nka repair center evaluated the device."change water trap" message was observed.Technician was unable to calibrate gas.The following part needed replacement: 1-part# cd-237p description: gas module.Service requested: evaluation and repair.Service performed: replaced 1-part# cd-237p description: gas module.Investigation conclusion: operator's manual revision j stated that gas calibration should be performed yearly for stable measuring accuracy.When the accuracy is not appropriate, perform gas calibration.The maintenance history for this device is unknown.Nka repair center evaluated the device and recommended replacing cd-237p gas unit sensor to resolve the issue.When repair services completed replacing the sensor, the issue was resolved.Device operated to manufacturer's specification.Device was put into service on (b)(6) 2012.The root cause of the sensor failure is unknown, however it operated as intended past the five-year warranty period for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AG-920RA
Common Device Name
MULTI-GAS UNIT
Manufacturer (Section D)
NIHON KOHDEN CORPORATION
1-31-4 nishiochia, shinjuku-ku
attn: shama mooman
tokyo, 161-8 560
JA  161-8560
MDR Report Key9307259
Report Number8030229-2019-00639
Device Sequence Number13919330
Product Code CCK
UDI-Device Identifier00841983103162
UDI-Public00841983103162
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K020046
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/17/2020
Report Date (Section F) 03/17/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberAG-920RA
Device Catalogue NumberAG-920RA
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/06/2019
Was the Report Sent to FDA? Yes
Distributor Facility Aware Date03/16/2020
Device Age86 MO
Event Location Hospital
Date Report to Manufacturer03/17/2020
Initial Date Received by Manufacturer 10/17/2019
Supplement Date Received by Manufacturer03/16/2020
Initial Report FDA Received Date11/11/2019
Supplement Report FDA Received Date03/17/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Date Report Sent to FDA03/17/2020
-
-