| Model Number |
MMT-723LNAS |
| Medical Device Problem Codes |
Device Difficult to Program or Calibrate (1496); Moisture or Humidity Problem (2986); Patient Device Interaction Problem (4001)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
12/28/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.No conclusion can be drawn at this time.We therefore consider this report complete to the best of our knowledge.Device received with unresponsive buttons due to moisture damage at keypad traces.No button error alarm noted.Traces of corrosion found at the electronic assembly and motor assembly per visual inspection.Unable to perform the displacement, operating currents due to unresponsive buttons.Device received with keypad overlay peeling.The asr exemption e2015043 was revoked on february 4, 2019.This event was previously reported as an asr.Additional information or analysis results regarding this event will be reported as an mdr.
|
| |
|
Event or Problem Description
|
|
The customer reported via phone call that the customer insulin pump had a button error alarm and moisture exposure.Customer¿s blood glucose was unknown at the time of incident.No troubleshooting was performed.Insulin pump will not be returned for analysis.
|
| |
|