• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. 530G INSULIN PUMP MMT-551NAB; ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. 530G INSULIN PUMP MMT-551NAB; ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND Back to Search Results
Model Number MMT-551NAB
Medical Device Problem Codes Display or Visual Feedback Problem (1184); Device Difficult to Program or Calibrate (1496); Power Problem (3010)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 10/01/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.No conclusion can be drawn at this time.We therefore consider this report complete to the best of our knowledge.Device received with no (act, up and escape) button response due to corroded keypad traces.No button error alarm during testing noted.Unable to perform all functional testing including the displacement, operating currents, unexpected restart error test, rewind, basic occlusion, occlusion, prime compromised force sensor system and excessive no delivery test or verify unexpected battery power loss anomaly, low battery alert and back light anomaly due to buttons not responding.The asr exemption e2015043 was revoked on february 4, 2019.This event was previously reported as an asr.Additional information or analysis results regarding this event will be reported as an mdr.
 
Event or Problem Description
The customer reported via phone that the customer had to press buttons more than the once to respond.Customer¿s blood glucose was unknown at the time of incident.The insulin pump will not be returned for analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
530G INSULIN PUMP MMT-551NAB
Common Device Name
ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
MDR Report Key9308146
Report Number2032227-2019-115940
Device Sequence Number2374418
Product Code OZO
UDI-Device Identifier00643169404250
UDI-Public(01)00643169404250
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 11/11/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberMMT-551NAB
Device Catalogue NumberMMT-551NAB
Device Lot NumberA3551NABJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/20/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/01/2019
Initial Report FDA Received Date11/12/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age62 YR
-
-