|
(b)(4).A complete analysis and testing of the insulin pump showed that it was functioning properly and passed all functional testing.After testing it was concluded that the device operated within specifications.Device was received with a minor scratched lcd window and a cracked reservoir tube lip.Device passed all functional tests, including the idle current measurement test, run current measurement test, self test, off no power test, unexpected restart error test, displacement test, basic occlusion test, occlusion test, prime/compromised force sensor system test, rewind test and excessive no delivery test.Device was received with normal operating currents.Device was monitored for several hours and no unexpected low battery alert noted during the testing.No motor error alarm noted.Motor passed motor test.The asr exemption e2015043 was revoked on february 4, 2019.This event was previously reported as an asr.Additional information or analysis results regarding this event will be reported as an mdr.
|