• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US PFC SIGMARP STB TB IN 6 12.5; SIGMA KNEE PRIMARY : KNEE TIBIAL INSERT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US PFC SIGMARP STB TB IN 6 12.5; SIGMA KNEE PRIMARY : KNEE TIBIAL INSERT Back to Search Results
Catalog Number 962162
Medical Device Problem Code Naturally Worn (2988)
Health Effect - Clinical Code No Code Available (3191)
Date of Event 10/16/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Patient was revised due to poly wear.Doi: (b)(6) 2007; dor: (b)(6) 2019 left knee.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary
=
> the device associated with this report was not received for examination.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PFC SIGMARP STB TB IN 6 12.5
Common Device Name
SIGMA KNEE PRIMARY : KNEE TIBIAL INSERT
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key9309818
Report Number1818910-2019-115597
Device Sequence Number467827
Product Code JWH
UDI-Device Identifier10603295241515
UDI-Public10603295241515
Combination Product (Y/N)N
Initial Reporter StateSC
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2007
Device Explanted Year2019
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/16/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number962162
Device Lot Number1942229
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 10/16/2019
Supplement Date Received by Manufacturer12/09/2019
Initial Report FDA Received Date11/12/2019
Supplement Report FDA Received Date12/10/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-