Model Number AB14W060040150
Medical Device Problem Code Burst Container or Vessel (1074)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 11/11/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Physician was attempting to use nanocross balloon catheter to treat the sfa.The device was prepped per ifu with no issues identified.It was reported that during inflation, the balloon burst at 8 atm.The angioplasty was achieved with balloon prior to burst.No other intervention to follow.There was no patient injury reported.
 
Additional Manufacturer Narrative
The type of burst is unknown.There were no balloon fragments and the balloon was attached to the catheter upon device removal from the patient.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
Product analysis: the nanocross dilatation catheter was returned to medtronic for evaluation within its shelf box.No ancillary device , cine, images or procedure notes were returned for evaluation.The device was removed from the packaging for additional analysis.The nanocross dilatation catheter was received with sanguine residue within the inflation portion of the y-manifold, sanguine residue/fluid within the balloon chamber, and the balloon chamber received in a post-inflation profile, (e.G.Not tightly wrapped or winged).A 10cc air filled syringe was attached to the proximal hub and the guidewire lumen was flushed: red fluid was observed exiting the distal tip.A 10cc water filled syringe was attached to the proximal hub luer lock and the guidewire lumen was flushed: a steady stream of red fluid was observed exiting the distal tip of the catheter.A 10cc water filled syringe was attached to the inflation lumen luer lock of the y-manifold.The balloon chamber was inflated to ap proximately 4 atm , and a leak was located in the balloon chamber on the proximal side.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NANOCROSS ELITE 0.14 OTW PTA DILATATION CATHETER
Common Device Name
CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL
Manufacturer (Section D)
COVIDIEN
4600 nathan lane north
plymouth MN 55442
MDR Report Key9310247
Report Number2183870-2019-00536
Device Sequence Number2360685
Product Code LIT
UDI-Device Identifier00763000110406
UDI-Public00763000110406
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
K132777
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 02/04/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/15/2022
Device Model NumberAB14W060040150
Device Catalogue NumberAB14W060040150
Device Lot NumberA883567
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/15/2019
Initial Date Received by Manufacturer 11/11/2019
Supplement Date Received by Manufacturer11/12/2019
02/03/2020
Initial Report FDA Received Date11/12/2019
Supplement Report FDA Received Date11/14/2019
02/04/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-