MAUDE Adverse Event Report: STRYKER ENDOSCOPY-SAN JOSE MICROFX OCD UNIVERSAL DRILL; BIT, DRILL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ENDOSCOPY-SAN JOSE MICROFX OCD UNIVERSAL DRILL; BIT, DRILL Back to Search Results
Catalog Number 234-200-200
Medical Device Problem Code Material Fragmentation (1261)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 10/17/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
 
Event or Problem Description
It was reported the tip broke during the procedure.The pieces were retrieved.
 
Additional Manufacturer Narrative
The product was not returned for investigation therefore the reported failure mode was not confirmed.The failure mode will be monitored for future reoccurrence.Alleged failure: tip break.Probable root cause: design: -drill too dull or dulls quickly, inefficient flute design, requires excessive force to advance.-drill not in alignment with guide.-drill allowed to extend too far out of guide.Process: -drill manufactured out of specification; -drill guide manufactured out of specification.Application: -excessive axial force on drill; -run drill in reverse; -reuse/resterilization of single-use device, or use of a single drill on multiple defects; -guide is translated and/or rotated during drilling; drill not in alignment with guide; -user does not use snap cap, drills too deep; -drill is started/stopped while in bone.The device manufacture date is not known.
 
Event or Problem Description
It was reported the tip broke during the procedure.The pieces were retrieved.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MICROFX OCD UNIVERSAL DRILL
Common Device Name
BIT, DRILL
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
MDR Report Key9310305
Report Number0002936485-2019-00504
Device Sequence Number16615274
Product Code HTW
UDI-Device Identifier37613327004091
UDI-Public37613327004091
Combination Product (Y/N)N
Initial Reporter CountryAU
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2019
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 02/18/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number234-200-200
Device Lot Number140461
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 10/17/2019
Supplement Date Received by Manufacturer10/17/2019
Initial Report FDA Received Date11/12/2019
Supplement Report FDA Received Date02/18/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age42 YR
Patient Weight90
-
-