PENUMBRA, INC. RUBY COIL; HCG, KRD Back to Search Results
Catalog Number RBY4C0305
Medical Device Problem Code Premature Separation (4045)
Health Effect - Clinical Codes No Consequences Or Impact To Patient (2199); Device Embedded In Tissue or Plaque (3165)
Date of Event 10/17/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The product was not returned for evaluation.Without the return of the device, the root cause of the problem cannot be determined.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
 
Event or Problem Description
The patient was undergoing a coil embolization procedure in the hepatic artery using ruby coils.During the procedure, the physician placed a ruby coil in the target vessel using a lantern delivery microcatheter (lantern).While advancing another ruby coil into the target vessel, the ruby coil was kicking the lantern kicked out of the aneurysm and into the proper hepatic artery; therefore, the ruby coil was retracted.However, while retracting, the ruby coil unintentionally detached, with part of the coil in the aneurysm and the other part in the lantern.The physician attempted to remove the lantern with aspiration to secure the detached ruby coil; however, the ruby coil ended up in the common hepatic artery and celiac trunk.A snare device was then used to remove the detached ruby coil; however, the coil was pushed by the flow of the vessel from the aneurysm into the proper hepatic artery and gastroduodenal artery (gda).The physician decided to end the procedure and leave the detached coil as a non-targeted embodiment in order to avoid further complications.There was no report of an adverse effect to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RUBY COIL
Common Device Name
HCG, KRD
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key9310733
Report Number3005168196-2019-02110
Device Sequence Number19292892
Product Code HCG
UDI-Device Identifier00814548018317
UDI-Public00814548018317
Combination Product (Y/N)Y
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K173614
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 10/17/2019
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberRBY4C0305
Device Lot NumberF92026
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/17/2019
Initial Report FDA Received Date11/12/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Date Report Sent to FDA01/01/2005
-
-