| Model Number |
401735 |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
10/21/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The orsiro drug-eluting stent system was selected for the treatment of a calcified and tortuous lesion.Before it was possible to deploy the orsiro the stent dislodged from the balloon.Therefore it was removed from the patients body with a threader.
|
| |
|
Additional Manufacturer Narrative
|
|
The returned product was subjected to a technical analysis and the corresponding production documentation was reviewed to establish whether a deviation from the manufacturing process could be the cause for this event.The technical investigation of the returned instrument revealed that the balloon is not well folded anymore.The stent was returned being deformed and partially put back onto the balloon.Stent imprints are visible on the balloon surface, indicating that the stent was initially crimped centered on the balloon.Review of the production documentation confirmed that the instrument was manufactured according to specifications and passed all inprocess and final inspections.As a part of final inspection every stent system under-goes visual inspection to ensure correct embedding and homogeneous crimping of the stent.Further, the stent outer diameter is verified to 100 percent and the stent retention force of a defined amount of samples is tested from every lot.Based on the conducted investigations, no manufacturing or material related root cause could be determined.
|
| |
|