• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. MOTOR DRIVE UNIT HAND CNTRL PWRMX EL SAW, POWERED, AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. MOTOR DRIVE UNIT HAND CNTRL PWRMX EL SAW, POWERED, AND ACCESSORIES Back to Search Results
Catalog Number 72200616
Device Problem Smoking (1585)
Patient Problem No Information (3190)
Event Type  Malfunction  
Event Description

It was reported that the dii controller had smoke coming out of the port. It is unknown whether the allegation was found during a surgical case. No harm has been stated. All available information has been disclosed. If additional information should become available, a supplemental report will be submitted accordingly.

 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameMOTOR DRIVE UNIT HAND CNTRL PWRMX EL
Type of DeviceSAW, POWERED, AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107 6512
Manufacturer (Section G)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107 6512
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key9311450
MDR Text Key166035419
Report Number1643264-2019-00797
Device Sequence Number1
Product Code HAB
Combination Product (Y/N)N
Reporter Country CodeUS
PMA/PMN NumberEXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type HEALTH PROFESSIONAL,USER FACI
Reporter Occupation OTHER HEALTH CARE PROFESSIONAL
Type of Report Initial,Followup
Report Date 12/23/2019
1 Device Was Involved in the Event
0 PatientS WERE Involved in the Event:
Date FDA Received11/12/2019
Is This A Product Problem Report? Yes
Device Operator HEALTH PROFESSIONAL
Device Catalogue Number72200616
Was Device Available For Evaluation? No
Is The Reporter A Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received12/17/2019
Was Device Evaluated By Manufacturer? Device Not Returned To Manufacturer
Is The Device Single Use? No
Is this a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unkown

-
-