| Model Number |
5400003412 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
11/01/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A follow up report will be filed once the quality investigation is complete.Awaiting device return.
|
| |
|
Event or Problem Description
|
|
It was reported that the blade broke inside the packaging.It was also reported that there were no delays and no adverse consequences as a result of this event.It was further reported that the procedure was completed successfully.
|
| |
|
Additional Manufacturer Narrative
|
|
H6: investigation results indicate that the blade broke due to a "crush"/ excessive force during distribution or storage.The quality investigation is complete.
|
| |
|
Event or Problem Description
|
|
It was reported that the blade broke inside the packaging.It was also reported that there were no delays and no adverse consequences as a result of this event.It was further reported that the procedure was completed successfully.
|
| |
|