U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIVANOVA DEUTSCHLAND CENTRIFUGAL PUMP 5 (CP5); CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIVANOVA DEUTSCHLAND CENTRIFUGAL PUMP 5 (CP5); CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 60-01-04
Device Problem Appropriate Term/Code Not Available (3191)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/16/2019
Event Type  malfunction  
Manufacturer Narrative
There was no patient involvement.Livanova (b)(4) manufactures the centrifugal pump 5 (cp5).The incident occurred in (b)(6).A livanova field service representative was dispatched to the facility to investigate the device and found the cp5 positioned horizontally not conformed with the ifu requirement.According to the technician it cannot be excluded that some liquid entered the holes at the lower side of drive unit causing the reported condition.A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
Livanova (b)(4) received a report that when turning on the main power and trying to turn on a centrifugal pump 5 (cp5), an error message was displayed and an alarm sounded."cold start" procedure (rebooting) powering off/on the device did not solve the problem.After the second attempt, another error message was displayed.It was also reported that a burning smell was coming from the drive unit once the power was turned on.There was no report of patient injury.
 
Event Description
See initial report.
 
Manufacturer Narrative
The device drive unit was returned to livanova deutschland for further investigation.The error could be reproduced.The defect has been caused by ingress of fluid.Several electrical and mechanical components were affected.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTRIFUGAL PUMP 5 (CP5)
Type of Device
CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich
MDR Report Key9314384
MDR Text Key195747411
Report Number9611109-2019-00891
Device Sequence Number1
Product Code DWA
Combination Product (y/n)N
PMA/PMN Number
K112225
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 12/17/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/13/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number60-01-04
Was the Report Sent to FDA? No
Date Manufacturer Received11/18/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-