The report is late due to technical issues we experienced with esg.
Event or Problem Description
During a dbs implantation neurosurgery, the neurosurgeon finished the microelectrode recoding phase and placed the implanted lead in place.The neurosurgeon noticed that the headstage drive was moving on its own and had moved in 2 mm.The neurosurgeon disconnected the headstage cable and decided to continue the neurosurgery by using the manual electrode moving option and the dbs electrode was successfully implanted.No adverse event reported.