• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOTRONIK AG, BUELACH, SWITZERLAND ORSIRO 3.5/26; CORONARY DRUG-ELUTING STENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOTRONIK AG, BUELACH, SWITZERLAND ORSIRO 3.5/26; CORONARY DRUG-ELUTING STENT Back to Search Results
Model Number 364509
Device Problem Deformation Due to Compressive Stress (2889)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/24/2019
Event Type  malfunction  
Event Description
An orsiro drug-eluting stent system was selected to treat a calcified lesion in a tortuous proximal lad.After deployment of the stent the wire looped and got caught on stent struts when pulling back whereby the stent became deformed.
 
Manufacturer Narrative
Neither the complaint instrument nor the angiographic material was provided for analysis.Therefore no technical investigation on the subject could be performed.The production documentation was reviewed to establish whether a deviation from the manufacturing process could be the cause for the reported event.Review of the production documentation confirmed that the instrument was manufactured according to specifications and passed all in-process and final inspections.Based on the conducted investigations, no manufacturing or material related root cause could be determined.It should be noted that the ifu warns to exercise care when crossing a newly deployed stent with a coronary guidewire.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ORSIRO 3.5/26
Type of Device
CORONARY DRUG-ELUTING STENT
Manufacturer (Section D)
BIOTRONIK AG, BUELACH, SWITZERLAND
ackerstrasse 6
buelach CH-81 80
CH  CH-8180
MDR Report Key9315257
MDR Text Key166627699
Report Number1028232-2019-05014
Device Sequence Number1
Product Code NIQ
Combination Product (y/n)N
PMA/PMN Number
P170030
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 11/11/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/13/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/19/2020
Device Model Number364509
Device Catalogue NumberSEE MODEL NO.
Device Lot Number07185326
Was Device Available for Evaluation? No
Date Manufacturer Received01/09/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Hospitalization;
-
-