On (b)(6) 2019, abbott point of care (apoc) was contacted by a customer regarding i-stat troponin cartridges that yielded a suspected discrepant troponin result of 0.64 on a (b)(6) year old female patient.There was no additional patient information at the time of this report.Return product is available for investigation.Method: i-stat, date: 10/24/2019, collected: 16:47, tested: 16:47, result: 0.64ng/ml.Lab, 10/24/2019, 17:44, 17:44, <0.01ng/ml.Lab, 10/24/2019, 20:02, 20:02, <0.01ng/ml.At this time there is no reason to suspect a malfunction exists.The reporting decision was based on the limited information available that suggests the product was not performing within the variability of the assay.The investigation is underway.Per i-stat system manual: art: 714258-00q.Reportable range: 0.00 - 50.00.Reference range: 0.00 - 0.03 (represents the 0 to 97.5% range of results).Reference range: 0.00 - 0.08 (represents the 0 to 99% range of results).
|