Product event summary: the balloon catheter, 2af283 with lot 85897, was returned and analyzed.Visual inspection of the catheter showed the device was intact with no apparent issue.Smart chip verification indicated the catheter was used for seventeen applications on the date of the event.The catheter passed the performance test as specification.During the inflation and ablation phase the kink was observed on guide wire lumen inside the balloon segment.The dissection showed the guide wire lumen was kinked inside the balloons around the heath shrink at.78 and 1.18 inches from the tip of the catheter.In conclusion, the reported issue was confirmed through testing.The balloon catheter failed the returned product inspection due to a guide wire lumen kink.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.If information is provided in the future, a supplemental report will be issued.
|