Model Number 9734680
Medical Device Problem Code Material Integrity Problem (2978)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 10/22/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
No patient information provided as no patient was involved in this concern.The probe was returned to the manufacture for evaluation.After functional testing and visual/physical examination the reported issue was confirmed.The returned probe had a bent tip, impact marks at the back end, and severe galling lines on the shaft that will not allow insertion into any mating tracker.Mechanical damage was observed.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Medtronic received information regarding a navigation system.It was reported outside of a procedure that there were three thoracic probe tips that were damaged and stuck into two navlock orange trackers.No was patient present at time of discovery.
 
Additional Manufacturer Narrative
Correction was made to the analysis of the returned probe had a bent tip, impact marks at the back end, and severe galling lines on the shaft that will not allow insertion into any mating tracker.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
THORACIC PROBE
Common Device Name
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC
826 coal creek circle
louisville CO 80027
MDR Report Key9315922
Report Number1723170-2019-05590
Device Sequence Number13550782
Product Code OLO
UDI-Device Identifier00763000054205
UDI-Public00763000054205
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K124004
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source user facility
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 12/06/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Model Number9734680
Device Catalogue Number9734680
Device Lot Number190417
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/30/2019
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/22/2019
Supplement Date Received by Manufacturer11/15/2019
Initial Report FDA Received Date11/13/2019
Supplement Report FDA Received Date12/06/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-