MAUDE Adverse Event Report: ABBOTT VASCULAR NC TREK RX; CORONARY DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR NC TREK RX; CORONARY DILATATION CATHETER Back to Search Results
Model Number 1012451-20
Medical Device Problem Codes Difficult to Insert (1316); Difficult to Remove (1528)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 10/16/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Exemption number e2019001.The device is expected to be returned for investigation.It has not yet been received.A follow-up report will be submitted with all additional relevant information.
 
Event or Problem Description
It was reported that the protective sheath of the 3.5x20 mm nc trek balloon dilatation catheter (bdc)was difficult to be removed.The bdc was flushed and prepped per usual but the sheath was very tight and came off after a few attempts.An attempt was made to insert a guide wire, but it was unsuccessful.The device never entered the patient.A new, same size nc trek bdc was used to complete the procedure.There was no adverse patient effect or significant delay in procedure.No additional information was provided.
 
Additional Manufacturer Narrative
Visual, dimensional and functional inspections were performed on the returned device.The reported difficulty inserting the guide wire could not be confirmed.The reported difficulty removing the protective sheath could not be tested since the protective sheath was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents and/or complaints from this lot.The investigation was unable to determine a conclusive cause for the reported difficulty removing the protective sheath; however, the reported difficulty inserting the guide wire appears to be related to circumstances of the procedure.There is no indication of a product quality issue with respects to the design, manufacture, or labeling of this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NC TREK RX
Common Device Name
CORONARY DILATATION CATHETER
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
MDR Report Key9316515
Report Number2024168-2019-13509
Device Sequence Number16737607
Product Code LOX
UDI-Device Identifier08717648151996
UDI-Public08717648151996
Combination Product (Y/N)N
Initial Reporter CountryTC
PMA/510(K) Number
K103153
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/17/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/31/2021
Device Model Number1012451-20
Device Catalogue Number1012451-20
Device Lot Number90129G1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/13/2019
Initial Date Received by Manufacturer 10/23/2019
Supplement Date Received by Manufacturer02/03/2020
Initial Report FDA Received Date11/13/2019
Supplement Report FDA Received Date02/17/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-