| Model Number |
MSERIES BI-PHASIC |
| Medical Device Problem Code |
Defibrillation/Stimulation Problem (1573)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that during biomed testing, the device displayed a "defib fault 72" message.Complainant indicated that there was no patient involvement in the reported malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
The customer's report was observed during initial testing; however after disassembling the device to determine root cause, the customer's report was no longer seen.The device was put through extensive testing without duplicating the report.Recommend to replace the high voltage module as a precaution; however the customer declined repair of the device.Analysis for reports of this type has not identified an increase in trend.
|
| |
|