• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR XIENCE SIERRA; DRUG ELUTING CORONARY STENT DELIVERY SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR XIENCE SIERRA; DRUG ELUTING CORONARY STENT DELIVERY SYSTEM Back to Search Results
Catalog Number 1550225-18
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Cardiac Enzyme Elevation (1838); Occlusion (1984)
Date of Event 10/21/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Exemption number e2019001.The stent remains in the patient.Investigation is not yet complete.A follow up report will be submitted with all additional relevant information.
 
Event or Problem Description
It was reported that a 2.25x18mm xience sierra was implanted in the circumflex coronary artery on (b)(6) 2019.On (b)(6) 2019, the patient was re-admitted with elevated troponin enzymes.Imaging found that there was no flow down the circumflex artery.The circumflex artery was occluded in the mid portion where the stent had been implanted.A balloon was advanced to reestablish flow.No additional treatment was performed as the collateral circulation was sufficient.There was no clinically significant delay in the procedure and no adverse patient effects.No additional information was provided.
 
Additional Manufacturer Narrative
Exemption number e2019001-permits numbering sequence to begin with 10000, to avoid duplication of report numbers due to process transition.There may be gaps in numbering for reports submitted during the transition period.The device was not returned for evaluation.The reported patient effect of occlusion is listed in the xience sierra, everolimus eluting coronary stent system, instructions for use, as a known patient effect of coronary stenting procedures.A review of the lot history record and complaint history of the reported device could not be conducted because the lot number was not provided.A conclusive cause for the reported patient effect, and the relationship to the product, if any, cannot be determined; however, the subsequent treatments appear to be related to the circumstances of the procedure.There is no indication of a product quality issue with respect to manufacture, design or labeling of the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XIENCE SIERRA
Common Device Name
DRUG ELUTING CORONARY STENT DELIVERY SYSTEM
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
MDR Report Key9317192
Report Number2024168-2019-13519
Device Sequence Number10914205
Product Code NIQ
UDI-Device Identifier08717648227028
UDI-Public08717648227028
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
P110019
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/05/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Operator of Device Health Professional
Device Catalogue Number1550225-18
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 10/21/2019
Supplement Date Received by Manufacturer11/13/2019
Initial Report FDA Received Date11/13/2019
Supplement Report FDA Received Date12/05/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age61 YR
-
-