SMITH & NEPHEW, INC. R3 20DEG 36MM TRIAL56MM; PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL
|
Back to Search Results |
|
| Catalog Number |
71367956 |
| Medical Device Problem Code |
Defective Component (2292)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
10/29/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during a thr procedure, the sharp edges of the trial cut the gloves of the surgeon.Plastic liner had gouges.No delay.Backup device available.No impact or injury reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The associated complaint device was returned and evaluated.A visual inspection confirms the trial has multiple scratches, burrs and deep gouges in the plastic.The device shows extreme signs of wear/usage.There is no lot number available for this device.We recommend that all re-usable instruments be routinely inspected for wear/damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.
|
| |
|