• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. REVERSE CURETTE 7MM; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. REVERSE CURETTE 7MM; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT Back to Search Results
Catalog Number 71369517
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 10/21/2019
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during surgery the reverse curette 7mm handle broke while it was being used inside the patient.Nothing got into the patient.There was no harm to the patient or the staff.There was no procedural delays to the case.A backup device was available.
 
Additional Manufacturer Narrative
The associated device was returned and evaluated.A visual inspection of the returned device confirmed the stated failure mode.The plastic handle on the device fractured in half and both pieces were returned.The device was manufactured in 2008 and exhibits signs of extensive wear / usage.A review of complaint history did not reveal additional complaints for the listed batch.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REVERSE CURETTE 7MM
Common Device Name
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks road
memphis TN 38116
MDR Report Key9318649
Report Number1020279-2019-04022
Device Sequence Number10902622
Product Code LXH
UDI-Device Identifier03596010430137
UDI-Public03596010430137
Combination Product (Y/N)N
Initial Reporter StateLA
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/25/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number71369517
Device Lot Number08DSY0016
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/08/2019
Initial Date Received by Manufacturer 10/21/2019
Supplement Date Received by Manufacturer10/21/2019
Initial Report FDA Received Date11/13/2019
Supplement Report FDA Received Date11/25/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-