| Model Number |
11607704 |
| Medical Device Problem Codes |
Stretched (1601); Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Patient information requested but not provided.Although requested, product has not been received.A follow up report will be submitted with failure investigation results should the device be received for evaluation.
|
| |
|
Event or Problem Description
|
|
It was reported that the silicone section of a 3 port iv administration set ballooned while in the an alaris infusion pump channel.There was no harm to patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information provided: d.10, & h.6.(device code).The customer¿s report of ballooning was confirmed.The set was visually inspected for kinks, incomplete bonding engagements, holes/tears in the tubing or damages to the components.Visual inspection of the set noted deformation at the top of the silicone pump segment.No other anomalies or evidence of damage were observed upon initial visual inspection.Examination under magnification of the silicone tubing segment walls were found to be concentric.Functional and pressure testing resulted in no ballooning in any clamped testing but has been replicated in unclamped testing when the tubing had been visually observed to be compromised.The root cause for the source of the excessive pressure is unknown.
|
| |
|
Event or Problem Description
|
|
It was reported that the silicone section of a (3) port iv administration set "ballooned", while in the alaris infusion pump channel.It was further confirmed that there was no patient harm or impact as a result of this event.
|
| |
|