MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-523RNAP; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-523RNAP; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR Back to Search Results
Model Number MMT-523RNAP
Device Problems Adverse Event Without Identified Device or Use Problem (2993); No Apparent Adverse Event (3189)
Patient Problems Congestive Heart Failure (1783); Hyperglycemia (1905); Pneumonia (2011)
Event Date 11/05/2019
Event Type  Injury  
Manufacturer Narrative
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event Description
The customer reported via phone call that they were hospitalized on (b)(6) 2019 due to hyperglycemia, pneumonia, and congestive heart failure.The customer's blood glucose level was 500 mg/dl at the time of hospitalization.The customer was unable to rewind the insulin pump and put insulin vial reservoir on the insulin pump.The customer was unable to complete a carelink upload.The customer was advised to discontinue use of the insulin pump and to revert to the back-up plan.The insulin pump will be returned for analysis.
 
Manufacturer Narrative
Device passed the functional test, including the displacement test, rewind, basic occlusion test, occlusion test, prime/a33 test, excessive no delivery test and delivery accuracy test at 0.08715 inches.The stop current and run current measurement tests are within specification.Device also passed self test, off no power alarm test and a21 error test.Device uploaded properly using carelink.Device had intermittent button response on the esc, act, up arrow and down arrow buttons due to flattened dome switches.No button error alarm noted.During visual inspection, the keypad connector on the lcd/b was found locked properly.Device had minor scratched display window, scratched case, cracked battery tube threads, cracked case at the reservoir tube window corner, cracked reservoir tube window, missing end cap sticker and cracked reservoir tube lip.The test p-cap and reservoir does lock in place in the reservoir compartment.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-523RNAP
Type of Device
PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer Contact
gerwin de graaff
ceiba norte ind. park #50 road
juncos 00777--386
8185464805
MDR Report Key9319748
MDR Text Key166323989
Report Number2032227-2019-116927
Device Sequence Number1
Product Code OYC
UDI-Device Identifier00643169513082
UDI-Public(01)00643169513082
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 01/27/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/13/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model NumberMMT-523RNAP
Device Catalogue NumberMMT-523RNAP
Device Lot NumberA1523RNAPJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/18/2019
Is the Reporter a Health Professional? No
Date Manufacturer Received01/23/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/15/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Treatment
FRN-UNK-RSVR, UNOMED INF SET
Patient Outcome(s) Hospitalization;
Patient Age74 YR
Patient Weight140
-
-