| Model Number |
MMT-523NAL |
| Medical Device Problem Code |
Device Difficult to Program or Calibrate (1496)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
12/12/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.No conclusion can be drawn at this time.We therefore consider this report complete to the best of our knowledge.Device received with intermittent button response due to corroded keypad traces.No button error alarm during testing.Device passed self test.The asr exemption e2015043 was revoked on february 4, 2019.This event was previously reported as an asr.Additional information or analysis results regarding this event will be reported as an mdr.
|
| |
|
Event or Problem Description
|
|
The customer¿s parent reported via phone call the insulin pump alarmed button error.The customer¿s blood glucose was unknown at the time of incident.Customer¿s parent stated that the insulin pump alarmed button error and unable to clear the alarm.No significant events leading to button error were observed.Button error troubleshooting was performed and confirmed that time is advancing.The customer¿s parent was advised to have the customer discontinue use of pump and revert to back-up plan.The insulin pump will be returned for analysis.
|
| |
|
Search Alerts/Recalls
|