• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK IRELAND LTD ZIMMON BILIARY STENT SET; FGE CATHETER, BILIARY, DIAGNOSTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK IRELAND LTD ZIMMON BILIARY STENT SET; FGE CATHETER, BILIARY, DIAGNOSTIC Back to Search Results
Model Number G25794
Medical Device Problem Code Material Deformation (2976)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 10/17/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device evaluation: 1 x zebd-7-10 of lot # c1626419 was returned to cirl for evaluation open in its original packaging.Lab evaluation: the device involved in the complaint was evaluated in the laboratory on 08th november 2019.A kink was observed at the 2nd & 5th portholes on the pigtail.A 0.035 inch wire guide would not pass the kink.Prior to distribution zebd-7-10 devices are subjected to a visual inspection and functional checks to ensure device integrity.These inspections and functional checks are outlined in internal procedures in place at cirl.A review of the manufacturing records for zebd-7-10 of lot number c1626419 did not reveal any discrepancies that could have contributed to this complaint issue.The review of relevant manufacturing records, confirms the failure mode has not previously occurred with the current lot number.Based on the information available to date, there is no evidence to suggest that there are any manufacturing issues associated with lot number c1626419.It should be noted that the instructions for use ((b)(4)) states the following: ¿visually inspect with particular attention to kinks, bends and breaks.If any abnormality is detected that would prohibit proper working condition, do not use.¿ ¿care must be exercised when straightening the pigtail curls in order to avoid kinking or breaking the stent.¿ a definitive root cause could not be determined as the circumstances of use cannot be replicated in the laboratory.A possible root cause could be attributed to the use of a non-recommended wire guide with the device.It may be noted the device label indicates a 0.035¿ wire guide for use with this device and all simulated use testing to date has been completed using a 0.035¿ wire guide.As the smaller diameter 0.025¿ wire guide occupies less space in the lumen of the pigtail stent it is likely that the extent of the curvature of the pigtail will be greater when a 0.025¿ wire guide is used.As per additional information received a 0.025" wire guide was used.Complaint is confirmed as the failure was verified in the laboratory.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.Complaints of this nature will continue to be monitored for potential emerging trends.
 
Event or Problem Description
Event is fda mdr reportable based on the device malfunction reporting precedence for this device family for the issue of ¿stent kinked/ bent'.The user advanced olympus/ visiglide wire guide 0.025" to b2 (bile duct).Then the user straightened the pig tail with the straightener (zebd-7-10/ lot c1626419) and attempted to advance it over the wire guide but failed.The user found a kink on the pig tail.Another zebd-7-10 was used instead and the procedure was completed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZIMMON BILIARY STENT SET
Common Device Name
FGE CATHETER, BILIARY, DIAGNOSTIC
Manufacturer (Section D)
COOK IRELAND LTD
o halloran road
limerick
MDR Report Key9322041
Report Number3001845648-2019-00593
Device Sequence Number18932204
Product Code FGE
UDI-Device Identifier00827002257948
UDI-Public(01)00827002257948(17)220626(10)C1626419
Combination Product (Y/N)N
PMA/510(K) Number
K851962
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 10/17/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/26/2022
Device Model NumberG25794
Device Catalogue NumberZEBD-7-10
Device Lot NumberC1626419
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/05/2019
Was the Report Sent to FDA? No
Distributor Facility Aware Date10/17/2019
Event Location Hospital
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/18/2019
Initial Report FDA Received Date11/14/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-