• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON INC. VCL CT BRD UD 27IN 4-0 S/A PS-2 PRM MP; SUTURE, ABSORBABLE, SYNTHETIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON INC. VCL CT BRD UD 27IN 4-0 S/A PS-2 PRM MP; SUTURE, ABSORBABLE, SYNTHETIC Back to Search Results
Catalog Number J426H
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 10/24/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.
 
Event or Problem Description
It was reported that a patient underwent a general surgery procedure on (b)(6) 2019 and suture was used.The tip of the needle broke off while suturing dense tissue.The doctor wasn't able to find the broken tip of the needle.It was not reported how the procedure was completed.There were no patient consequences reported.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).A manufacturing record evaluation was performed for the finished device pj5576 batch number, and no non-conformances were identified.A suture needle was submitted for evaluation.A fracture was observed at the tip of the sample b.One side of needle was received, the mating fracture surface was not provided for this evaluation.A scanning electron microscope (sem) was used to examine the fracture surface and surrounding area of the needle.The fracture surface was examined in multiple locations in order to determine the fracture mode.The evaluation revealed the fracture was composed of microvoid coalescence, which is evidence of a ductile overload failure.Mechanical damage observed coincidental to the fracture provides additional evidence that the failure was induced by mechanical deformation leading to ductile overload.This was a ductile fracture.The manufacturing records were reviewed, and the manufacturing/ packaging criteria were met prior to the release of this batch.The evidence of this examination indicates that the breakage occurred at the tip of the needle during use due to tensile overload.There is no evidence of any material flaw or defect that would cause premature failure.In order to avoid this kind of damage grasp the needle in an area one-third (1/3) to one-half (1/2) of the distance from the attachment end to the point.Additional device sample a reported in mw 2210968-2019-89930.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VCL CT BRD UD 27IN 4-0 S/A PS-2 PRM MP
Common Device Name
SUTURE, ABSORBABLE, SYNTHETIC
Manufacturer (Section D)
ETHICON INC.
p.o. box 151, route 22 west
somerville NJ 08876
MDR Report Key9324039
Report Number2210968-2019-89931
Device Sequence Number13915636
Product Code GAM
UDI-Device Identifier10705031039773
UDI-Public10705031039773
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K022269
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/24/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/31/2024
Device Catalogue NumberJ426H
Device Lot NumberPJ5576
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/05/2019
Initial Date Received by Manufacturer 10/24/2019
Supplement Date Received by Manufacturer12/05/2019
Initial Report FDA Received Date11/14/2019
Supplement Report FDA Received Date12/13/2019
Patient Sequence Number1
-
-