| Model Number |
OPES 90, ASPIRATING ABLATOR, TOOTHBRUSH |
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
09/30/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.
|
| |
|
Event or Problem Description
|
|
It was reported that a shoulder arthroscopy was performed with an arthrex device.The head of the probe got deformed and detached by the effect of heat.According to the surgeon the detached fragment was not retrieved.No further information received at the moment.Update 25-oct-2019: after the surgery it was found that the probe had a broken edge on the head and a small piece was missing, the joint was well washed and searched for the missing piece, but was not found.
|
| |
|
Additional Manufacturer Narrative
|
|
Complaint confirmed.Visual evaluation revealed a portion of the ceramic head broken-off at the top left of the ceramic head facing the electrodes.Also, heat damaged was noticed at the top of the head, as a result of the broken piece of the ceramic.A most likely cause for this event to occur is by mechanical damage to device such as hitting the device with another device, prying/leveraging the device ,and/or running the device longer than intended during use.
|
| |
|