ENCORE MEDICAL L.P. X-ALT HIP; LINER, ACETABULAR, HOODED, NEUTRAL, 10°, HXL/FMP, 28MM ID, MP5
|
Back to Search Results |
|
| Model Number |
932-28-248 |
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
10/17/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The reason for this revision surgery was reported as worn poly.The previous surgery and the surgery detailed in this event occurred 9 years and 6 months apart.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The devices were disposed of at the hospital and not made available to djo surgical for examination.A review of the device history records (dhr) show that the reported components used in the previous surgery, when released for use, met design and manufacturing requirements.There were no non-conforming material reports (ncmr) associated with the products that may have contributed to the reported event.The devices were verified to have gone through an acceptable sterilization process and were within its expiration date at the time of the previous surgery.Customer complaint history of the reported devices showed no present trends or on-going issues that are needing a review.The root cause of this complaint was a revision surgery due to worn poly.There were no findings during this evaluation that indicate the reported devices were defective.No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the event.There are multiple factors that may contribute to an event that are outside the control of djo surgical such as inadequate soft tissue support, poor bone density, excessive range of motion, patient bone deterioration, degenerative bone disease, patient activities or trauma.There are no indications of a product or process issue affecting implant safety or effectiveness.
|
| |
|
Event or Problem Description
|
|
Revision surgery - the poly was worn in the patient, surgeon removed the femoral head and the femoral shell and liner.He replaced the acetabul shell and liner with biomet duel pivot system and replaced the femoral head with our 28mm delta option head and shell.
|
| |
|
Search Alerts/Recalls
|
|
|