Catalog Number 5011222
Medical Device Problem Code Material Deformation (2976)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 10/18/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Neither the device nor films of applicable imaging studies were returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that the patient underwent posterior lumbar interbody fusion (plif) at l5/s1 due to lumbar canal stenosis.Intra-op, when the inserter which was attached to the cage was inserted into the intervertebral space with a hammer (before rotating), it was found that the screw of the inserter had loosened.The screw was turned again, but the cage could not be fixed onto the inserter at all.There were damages and deformation found post-operatively on the cage that must had been generated during the removal.So the cage was once removed and another cage was attached to the inserter.The intervertebral disc space was narrow and it seems hammering was more than usual.As reported in the overall procedure was delayed by less than 60 mins.There was no patient complication reported as a result of this event.
 
Additional Manufacturer Narrative
H6: product analysis results: optical and microscopic examination of the implant returned for analysis identified deformation at the inserter mating face and threads are damaged.This is consistent with significant impact during attempted implantation.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAPSTONE CONTROL PTC SPINAL SYSTEM
Common Device Name
INTERVERTEBRAL FUSION DEVICE WITH BONE GRAFT, LUMBAR
Manufacturer (Section D)
MDT PUERTO RICO OPERATIONS CO, MED REL
road 909, km. 0.4., barrio mar
humacao PR 00792
MDR Report Key9325787
Report Number2647346-2019-00021
Device Sequence Number2498258
Product Code MAX
UDI-Device Identifier00643169677524
UDI-Public00643169677524
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K171107
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/20/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number5011222
Device Lot NumberH5530440
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/27/2020
Initial Date Received by Manufacturer 10/18/2019
Supplement Date Received by Manufacturer04/14/2020
Initial Report FDA Received Date11/15/2019
Supplement Report FDA Received Date04/20/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age62 YR
-
-