SMITH & NEPHEW, INC. SVCE REPL LIGHT SOURCE 500XL XENON; IMAGE, ILLUMINATION, FIBEROPTIC, FOR ENDOSCOPE Back to Search Results
Catalog Number 72200568S
Device Problem Overheating of Device (1437)
Patient Problem No Patient Involvement (2645)
Event Date 10/22/2019
Event Type  malfunction  
Event Description
It was reported that the light source was too hot.No case was involved.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Manufacturer Narrative
The reported device, intended for use in treatment, was received for evaluation.A visual inspection was performed on the exterior of product and no physical damage was observed.There was a relationship found between the returned device and the reported incident.The complaint of an e-12 errors was confirmed.E12 is the error code for overheating.The cooling fan for the ballast/power supply shuts down when unit is powered up causing overheating and errors.Power supply was also electrically damaged due to fan malfunction.Fan malfunction is caused by a defective mcu pcb.Unit passes functional testing and fans perform as expected with a known good mcu pcb and power supply/ballast installed.The complaint investigation has concluded the cause of the failure to be a defective electronic component on the mcu pcb.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SVCE REPL LIGHT SOURCE 500XL XENON
Type of Device
IMAGE, ILLUMINATION, FIBEROPTIC, FOR ENDOSCOPE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107 6512
MDR Report Key9329054
MDR Text Key177224200
Report Number1643264-2019-00803
Device Sequence Number1
Product Code FFS
Combination Product (y/n)N
PMA/PMN Number
K994084
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 12/19/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/15/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number72200568S
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/06/2019
Date Manufacturer Received12/17/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-