| Catalog Number |
ST18JN60NSZ165N |
| Medical Device Problem Code |
Inflation Problem (1310)
|
| Health Effect - Clinical Code |
Increased Respiratory Rate (2486)
|
| Date of Event |
10/09/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Information was received indicating that immediately following placement of a smiths medical portex bivona custom flextend¿ tts¿ tracheostomy tube, the cuff was observed to not be inflating properly.The patient was reported to not be getting the appropriate tidal volume needed and their respiratory rate was 30-35 respirations a minute.Subsequently, a trach tube change out was performed.The patient recovered with no further adverse patient effects.
|
| |
|
Additional Manufacturer Narrative
|
|
One custom 6.0mm id fixed tracheostomy tube with swivel connector and tts cuff was returned for analysis.The inflation valve was found to be pushed inside the balloon upon visual inspection.The inflation valve was reseated to its normal position, a syringe was used to inflate the cuff; cuff remained inflated.The air was then removed, and the cuff was leak tested.Based on the evidence, there was no fault found as the complaint was not confirmed.
|
| |
|