DEPUY IRELAND - 9616671 ACTIS BROACH SZ 1; BROACHES Back to Search Results
Catalog Number 201001010
Device Problems Naturally Worn (2988); Scratched Material (3020)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 09/17/2019
Event Type  malfunction  
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: examination of the returned instrument confirmed the complaint.Examination of the returned broach found the teeth exhibited light wear.Circular scratching was found on the post of the broach.The root cause is attributed to wear and tear from normal use and servicing.Depuy considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot: null.Device history batch: null.Device history review: null.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that the broaches are showing signs of wear.No surgical delay.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint # (b)(4).Examination of the returned instrument confirmed the complaint.Examination of the returned broach found the teeth exhibited light wear.Circular scratching was found on the post of the broach.The root cause is attributed to wear and tear from normal use and servicing.Depuy considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACTIS BROACH SZ 1
Type of Device
BROACHES
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
MDR Report Key9329788
MDR Text Key186725345
Report Number1818910-2019-116494
Device Sequence Number1
Product Code KWL
UDI-Device Identifier10603295393634
UDI-Public10603295393634
Combination Product (y/n)N
PMA/PMN Number
K150862
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 09/17/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/15/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number201001010
Device Lot NumberPG257838
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/01/2019
Date Manufacturer Received11/15/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age66 YR
-
-