Catalog Number 201001090
Device Problem Naturally Worn (2988)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 09/17/2019
Event Type  malfunction  
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: examination of the returned instrument confirmed the complaint.Examination of the returned broach found the teeth exhibited light wear.Circular scratching was found on the post of the broach.The root cause is attributed to wear and tear from normal use and servicing.Depuy considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot: null.Device history batch: null.Device history review: null.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that the broaches are showing signs of wear.No surgical delay.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACTIS BROACH SZ 9
Type of Device
BROACHES
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic dr.
warsaw IN 46581 0988
Manufacturer Contact
kara ditty-bovard
700 orthopaedic dr.
warsaw, IN 46581-0988
6107428552
MDR Report Key9329844
MDR Text Key186725297
Report Number1818910-2019-116503
Device Sequence Number1
Product Code KWL
UDI-Device Identifier10603295393719
UDI-Public10603295393719
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K150862
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 09/17/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number201001090
Device Lot NumberPG261646
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/01/2019
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/07/2019
Initial Date FDA Received11/15/2019
Date Device Manufactured12/13/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age65 YR
-
-