| Model Number |
B2040-080 |
| Medical Device Problem Codes |
Inflation Problem (1310); Leak/Splash (1354)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
10/30/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device is expected to be returned for investigation.It has not yet been received.A follow-up report will be submitted with all additional relevant information.Exemption number e2019001.
|
| |
|
Event or Problem Description
|
|
It was reported that the procedure was performed to treat a moderately calcified, moderately tortuous superficial femoral artery that was 80% stenosed.The 4.0x80mm armada 35 balloon dilatation catheter was pressurized to nominal pressure.The balloon only partially inflated and a leak was noted from the hub.A new 4.0x60mm armada 35 was used to successfully complete the procedure.There were no adverse patient effects and no clinically significant delay in the procedure.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned for evaluation.The leak and inflation issue was confirmed.A review of the lot history record identified no nonconforming material records (ncmr).Additionally, there have been no other incidents and/or complaints reported for this lot.The investigation determined that the reported hub leak appears to be related to a potential product quality issue.The issue is being addressed per internal operating procedures.Abbott vascular will continue to trend the performance of these devices.
|
| |
|