• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR ARMADA 35 PTA CATHETER; PERIPHERAL DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR ARMADA 35 PTA CATHETER; PERIPHERAL DILATATION CATHETER Back to Search Results
Model Number B2040-080
Medical Device Problem Codes Inflation Problem (1310); Leak/Splash (1354)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 10/30/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device is expected to be returned for investigation.It has not yet been received.A follow-up report will be submitted with all additional relevant information.Exemption number e2019001.
 
Event or Problem Description
It was reported that the procedure was performed to treat a moderately calcified, moderately tortuous superficial femoral artery that was 80% stenosed.The 4.0x80mm armada 35 balloon dilatation catheter was pressurized to nominal pressure.The balloon only partially inflated and a leak was noted from the hub.A new 4.0x60mm armada 35 was used to successfully complete the procedure.There were no adverse patient effects and no clinically significant delay in the procedure.No additional information was provided.
 
Additional Manufacturer Narrative
The device was returned for evaluation.The leak and inflation issue was confirmed.A review of the lot history record identified no nonconforming material records (ncmr).Additionally, there have been no other incidents and/or complaints reported for this lot.The investigation determined that the reported hub leak appears to be related to a potential product quality issue.The issue is being addressed per internal operating procedures.Abbott vascular will continue to trend the performance of these devices.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARMADA 35 PTA CATHETER
Common Device Name
PERIPHERAL DILATATION CATHETER
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
MDR Report Key9329855
Report Number2024168-2019-13626
Device Sequence Number398621
Product Code LIT
UDI-Device Identifier08717648155109
UDI-Public08717648155109
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K111899
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/14/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/31/2022
Device Model NumberB2040-080
Device Catalogue NumberB2040-080
Device Lot Number90212G1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/14/2020
Initial Date Received by Manufacturer 10/30/2019
Supplement Date Received by Manufacturer02/28/2020
Initial Report FDA Received Date11/15/2019
Supplement Report FDA Received Date03/14/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
COMMAND GUIDE WIRE
Patient Age80 YR
Patient Weight76
-
-