• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOMET MICROFIXATION ELEVATOR #0; ELEVATOR, SURGICAL, DENTAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET MICROFIXATION ELEVATOR #0; ELEVATOR, SURGICAL, DENTAL Back to Search Results
Model Number N/A
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Zimmer biomet complaint (b)(4).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
It was reported that an instrument fractured during a dental procedure.The fractured piece of the instrument was removed from the patient by suction and the procedure was completed with a back-up instrument.No adverse events were reported as a result of this malfunction.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The complaint is confirmed.The complaint was opened because the facility reported the tip of the instrument broke off.The elevator #0 (09-0258, lot# 032613c13) was returned for investigation.Visual evaluation showed signs of use as there was minor scratching on the body of the elevator and the tip had fractured off.The non-conformance database was reviewed for this product, no non-conformances were found.Complaint history for 09-0258, lot# 032613c13 was reviewed and there is a complaint rate of 0.1%, which is no greater than the occurrence listed in the application fmea.The most likely underlying cause of the complaint is that excessive force was used, beyond what the instrument was designed to encounter.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.The following fields were updated: b4 date of this report b5 describe event or problem d4 lot number d10 device availability g4 date received by manufacturer g7 type of report h2 follow up type h3 device evaluated by manufacturer h6 method code h6 results code h6 conclusions code h10 additional narratives/data.
 
Event or Problem Description
This follow-up report is being submitted to relay additional information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELEVATOR #0
Common Device Name
ELEVATOR, SURGICAL, DENTAL
Manufacturer (Section D)
BIOMET MICROFIXATION
1520 tradeport drive
jacksonville FL 32218
MDR Report Key9330484
Report Number0001032347-2019-00489
Device Sequence Number2243703
Product Code EMJ
UDI-Device Identifier00841036028190
UDI-Public00841036028190
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
N/A
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Dentist
Type of Report Initial,Followup
Report Date (Section B) 01/28/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number09-0258
Device Lot Number032613C13
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/13/2020
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 10/17/2019
Supplement Date Received by Manufacturer01/13/2020
Initial Report FDA Received Date11/15/2019
Supplement Report FDA Received Date01/28/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
-
-