• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER INSTRUMENTS-KALAMAZOO NAVLOCK, UNIVERSAL TRACKER ADAPTER FOR INSTRUMENTS FROM 13 TO 20MM; NEUROLOGICAL STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER INSTRUMENTS-KALAMAZOO NAVLOCK, UNIVERSAL TRACKER ADAPTER FOR INSTRUMENTS FROM 13 TO 20MM; NEUROLOGICAL STEREOTAXIC INSTRUMENT Back to Search Results
Model Number 6000-999-004
Medical Device Problem Codes Material Fragmentation (1261); Material Deformation (2976)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 10/21/2019
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during set up for a procedure the device was found stripped and broken.The device being fragmented poses a risk of a small component being lost in the surgical site.No medical intervention and no adverse consequences were reported with this event.As this event occurred during setup at the user facility, there was no patient involvement and no delay to a surgical procedure.
 
Additional Manufacturer Narrative
Device evaluation: follow-up report submitted to document device evaluation results.
 
Event or Problem Description
It was reported that during set up for a procedure the device was found stripped and broken.The device being fragmented poses a risk of a small component being lost in the surgical site.No medical intervention and no adverse consequences were reported with this event.As this event occurred during setup at the user facility, there was no patient involvement and no delay to a surgical procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NAVLOCK, UNIVERSAL TRACKER ADAPTER FOR INSTRUMENTS FROM 13 TO 20MM
Common Device Name
NEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
STRYKER INSTRUMENTS-KALAMAZOO
4100 east milham avenue
kalamazoo MI 49001
MDR Report Key9331231
Report Number0001811755-2019-03764
Device Sequence Number16714890
Product Code HAW
UDI-Device Identifier04546540391926
UDI-Public04546540391926
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/26/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number6000-999-004
Device Catalogue Number6000-999-004
Device Lot NumberC/10/2018
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/25/2019
Initial Date Received by Manufacturer 10/21/2019
Supplement Date Received by Manufacturer02/26/2020
Initial Report FDA Received Date11/15/2019
Supplement Report FDA Received Date02/26/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-