• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION 2CM PERIPHERAL CUTTING BALLOON CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINA

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION 2CM PERIPHERAL CUTTING BALLOON CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINA Back to Search Results
Model Number 24634
Device Problem Material Rupture (1546)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/29/2019
Event Type  malfunction  
Event Description
It was reported that balloon rupture occurred. The 99% stenosed target lesion was located in a moderately tortuous and severely calcified common femoral artery to superficial femoral artery. A 4. 00mm/1. 5cm/140cm small peripheral cutting balloon and a 6. 00mm/2. 0cm/135cm peripheral cutting balloon were selected for use. During the procedure, a non bsc guidewire was advanced; however, it failed to cross the lesion. After several attempts, the wire eventually crossed the lesion and then dilation was performed with the 4mm small peripheral cutting balloon. However, during second inflation, it was noted that the balloon ruptured at 6atm. Subsequently, the 6mm peripheral cutting balloon was used; however, during second inflation, the balloon ruptured at 10atm. The devices were simply removed from the patient's body. The procedure was completed with a different device. No patient complications were reported and the patient was good post procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name2CM PERIPHERAL CUTTING BALLOON
Type of DeviceCATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINA
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key9332365
MDR Text Key173347211
Report Number2134265-2019-13940
Device Sequence Number1
Product Code LIT
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation
Type of Report Initial
Report Date 11/15/2019
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received11/15/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Expiration Date01/27/2021
Device Model Number24634
Device Catalogue Number24634
Device Lot Number0023252457
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received10/29/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/28/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-