| Model Number |
PCF-H180AI |
| Medical Device Problem Code |
Microbial Contamination of Device (2303)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The subject device in this report has not been returned to omsc for evaluation.The exact cause of the reported event could not be conclusively determined at this time.If additional information becomes available, this report will be supplemented.
|
| |
|
Event or Problem Description
|
|
Olympus medical systems corp.(omsc) was informed that as a result of routine microbiological testing by the user facility on the sample collected from the subject device tested positive for enterobacteriaceae and pseudomonas aeruginosa (greater than 100 cfu).The device had been reprocessed with a non-olympus automated endoscope reprocessor, soluscope 4, using peracetic acid.There was no report of infection associated with this report.
|
| |
|
Additional Manufacturer Narrative
|
|
This supplemental report is being submitted to provide additional information.The subject device has not been returned to omsc but was returned to olympus (b)(4).(b)(4) sent the device to a third party laboratory for microbiological testing.As a result of the testing, no microbe was detected from the subject device.The testing result cleared the french guideline.The exact cause of the reported event could not be conclusively determined at this time.If additional information becomes available, this report will be supplemented.
|
| |
|
Additional Manufacturer Narrative
|
|
This supplemental report is being submitted to provide additional information.Omsc reviewed the manufacturing history (dhr) of the subject device and confirmed no irregularity.The exact cause of the reported event could not be conclusively determined.If additional information becomes available, this report will be supplemented.
|
| |
|